BS ISO 21501-2:2019 Determination of particle size distribution. Single particle light interaction methods Light scattering liquid-borne particle counter

BS ISO 21501-2:2019

Determination of particle size distribution. Single particle light interaction methods Light scattering liquid-borne particle counter

Formát
Dostupnost
Cena a měna
Anglicky Zabezpečené PDF
K okamžitému stažení
6240 Kč
Anglicky Tisk
Skladem
6240 Kč
Označení normy:BS ISO 21501-2:2019
Počet stran:30
Vydáno:2019-11-27
ISBN:978 0 539 02807 2
Status:Standard
Popis

BS ISO 21501-2:2019


This standard BS ISO 21501-2:2019 Determination of particle size distribution. Single particle light interaction methods is classified in these ICS categories:
  • 19.120 Particle size analysis. Sieving

This document describes a calibration and verification method for a light scattering liquid-borne particle counter (LSLPC), which is used to measure the size and particle number concentration of particles suspended in liquid. The light scattering method described in this document is based on single particle measurements. The typical size range of particles measured by this method is between 0,1 µm and 10 µm in particle size.

The method is applicable to instruments used for the evaluation of the cleanliness of pure water and chemicals, as well as the measurement of number and size distribution of particles in various liquids. The measured particle size using the LSLPC depends on the refractive index of particles and medium; therefore, the measured particle size is equivalent to the calibration particles in pure water.

The following are within the scope of this document:

  • size setting error;

  • counting efficiency;

  • size resolution;

  • false count;

  • maximum particle number concentration;

  • sampling flow rate error;

  • sampling time error;

  • sampling volume error;

  • calibration interval;

  • reporting results from test and calibration.